Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

Scientific Reports volume  12, Article number: 15855 (2022 ) Cite this article

This paper presents the gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4 \(\times\) 4) with \(\hbox {LaBr}_3\) (Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newly developed array based on MAPD-3NM-II sensors proved \(\sim\)  22% enhancement in energy resolution in comparison to the former MAPD-3NM-I based array. The energy resolution of 662 keV gamma-rays measured by MAPD-3NM-II was 3.3% while clearly surpassing 4.25% resolution of MAPD-3NM-I predecessor. The enhancement is related to the high PDE of the new MAPD-3NM-II. Obtained results show that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD.

Recently, unmanned aerial vehicle (UAV) technology has been used widely in different fields from the army to security1,2,3,4,5,6,7,8. The use of drones in these areas makes it possible to reduce the risk to human life, to control the situation very quickly and accurately1. Drones are now commonly being used for environmental radiation monitoring and radiological mapping following a nuclear or radiological emergency, such as at Fukushima in Japan and the Chernobyl nuclear power plant in Ukraine4,5,6,7,8. The main requirements on radiation monitoring devices used in drone technology are low power consumption, light mass, humidity resistance, resistance to vibration and shock, and so on8. Conventional scintillation detectors based on vacuum photomultiplier tubes (PMT) do not meet these specific operational conditions. Therefore, scintillation detectors based on new photosensors are quite promising candidates that meet the requirements much better, and focusing on their further development is in high demand9,10,11,12,13,14,15,16,17,18,19,20,21,22. Many research teams focus on the development of scintillation detectors based on micro pixel avalanche photodiodes (MAPD) or silicon avalanche photomultipliers (SiPM)15,16,17,18,19,20,21,22,23,24,25,26. The parameters of latest SiPM generations allows them to become applicable as effective sensor element replacements of PMT in scintillation detectors. There are many works demonstrating the gamma-ray detection performance of scintillation detectors based on SiPM23,24,25,26,27. The main requirements for SiPM in this application are high pixel density and photon detection efficiency (PDE) which play a key role in the linearity and energy resolution of the entire detector17,21,23,24,25,26.

In this work, we demonstrate the gamma-ray detection performance of the new MAPD-3NM-II type SiPM sensor array assembled with \(\hbox {LaBr}_3\) (Ce) scintillators.

The cubic \(\hbox {LaBr}_3\) (Ce) crystal (15 \(\times\) 15 \(\times\) 30 \(\hbox {mm}^3\) ) used in measurements was produced by Epic Crystal28. \(\hbox {LaBr}_3\) (Ce) crystal is hermetically sealed in a thin aluminum package with a 3 mm light guide on one face (Fig. 1 left). The scintillator was mounted on the MAPD-3NM-II type SiPM array with optical grease.

Left—photo of the \(\hbox {LaBr}_3\) (Ce) scintillator (15 \(\times\) 15 \(\times\) \(30\,\hbox {mm}^3\) ), right—photo of MAPD-3NM-II type SiPM array embedded in the protective casing.

Parameters of the used scintillator are given in Table 1.

The MAPD-3NM-II is used as the photon readout of the \(\hbox {LaBr}_3\) (Ce) scintillator. Due to the large size of \(\hbox {LaBr}_3\) :Ce scintillator, 16 SiPMs of the MAPD-3NM-II type are combined in parallel to obtain a large sensitive area. MAPD-3NM-II with buried pixel structure manufactured together with Zecotek Company in 2020 (Fig. 1 right)23,29. The array is made up of sixteen 3.7 \(\times\) 3.7 \(\hbox {mm}^2\) MAPD elements and each element consists of 61000 pixels. A positive voltage was applied to the cathode, and the common anode was used as an output. The gap between two MAPD elements is 0.1 mm, as designed in the PCB layout. The total area covered by the 16 MAPDs is 17 \(\times\) 17 \(\hbox {mm}^2\) . The technical specifications and parameters of the MAPD-3NM-II array are listed in Table 2.

The operation and breakdown voltage of MAPD array is defined from the relative derivative of the I–V curve. The reverse current–voltage (I–V) characteristic is performed using a Keithley 6487 Picoammeter/Voltage Source. The breakdown voltage (\(\hbox {U}_{{br}}\) ) was 51.6 V and the full working voltage range of the MAPD-3NM-II array was 54–56 V.

The total capacitance of the array was \(\sim\) 2.5 nF. Compared to the previous MAPD-3NM-I element, the following parameters of the new MAPD-3NM-II was improved thanks to optimization of the structure: the operating voltage decreased by 24%, the dark current decreased more than 2 times at the same gain, the capacitance decreased by 23%, the gain increased 2 times, and the PDE increased more than 30%23.

This experiment was performed with various radio-isotope standard sources. The gamma-ray detection performance of the detector is tested with \(^{241}\hbox {Am}\) , \(^{133}\hbox {Ba}\) , \(^{137}\hbox {Cs}\) , and \(^{60}\hbox {Co}\) point radioactive sources. The parameters of radioactive sources are shown in Table 3. The radioactive sources were placed on the top of the \(\hbox {LaBr}_3\) (Ce) scintillator inside the dark box.

Gamma-ray measurements were carried out in the SPECTRIG MAPD module, where the temperature was \(23.5\, ^{\circ }\hbox {C}\) . The detailed information about the SPECTRIG MAPD module was presented in22. During measurement with SPECTRIG MAPD following parameters were selected: gate width – 227 nsec, variable gain – 1 dB (for M MAPD-3NM-II) and 10 dB (for MAPD-3NM-I), threshold – 6 mV, applied voltage – 54.5 V, 55.8 V (for MAPD-3NM-II) and 76.5 V (for MAPD-3NM-I). High-energy gamma rays (e.g. E >2 MeV) did not appear in the spectrum due to saturation of the ADC output in the case of measurements with MAPD-3NM-II at different voltages. A measurement time of 5 minutes for each cycle was selected.

Figure 2 shows the spectrum of the mixed radioisotope sources containing (\(^{241}\hbox {Am}\) , \(^{133}\hbox {Ba}\) and \(^{137}\hbox {Cs}\) ).

Energy spectra of mixed radioisotope sources containing (\(^{241}\hbox {Am}\) , \(^{133}\hbox {Ba}\) and \(^{137}\hbox {Cs}\) ) at an applied voltage of 55.8 V for \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II.

The high photon detection efficiency of MAPD-3NM-II and the high light output of the \(\hbox {LaBr}_3\) scintillator made it possible to obtain a good resolution. All this made it possible to separate the photopeak of the gamma-rays. It is obvious that the newly developed MAPD detector allows detecting most of the gamma lines emitted from \(^{133}\hbox {Ba}\) radioisotope. This kind of detector allowed to separate close gamma lines, e.g. 59.6 keV (\(^{241}\hbox {Am}\) ) from 81 keV (difference \(\sim\) 21 keV), 276.3 keV from 302.6 keV (difference \(\sim\) 26 keV) and 356 keV from 383.8 keV (difference \(\sim\) 28 keV). The energy resolution was evaluated by fitting the photopeak of the \(^{137}\hbox {Cs}\) energy spectrum with a Gaussian function. The energy resolution of 662 keV gamma rays was 3.3%.

In the case of measurements of \(^{60}\hbox {Co}\) with MAPD-3NM-II at 55.8 V voltage, the output of the ADC of SPECTRIG MAPD module was saturated (even when a variable gain was set 1 dB, i.e. minimal) that is why high-energy gamma-rays (e.g. E > 1 MeV) was not seen at this voltage. This disadvantage was avoided by decreasing the operation voltage to 54.5 V. However, as a side-effect of this approach, the energy resolution of the detector gets reduced but detection of high-energy gamma-rays (e.g. E > 1 MeV) is possible.

A spectrum of mixed radioactive sources containing \(^{241}\hbox {Am}\) + \(^{133}\hbox {Ba}\) + \(^{137}\hbox {Cs}\) + \(^{60}\hbox {Co}\) is depicted in Fig. 3 (left).

Energy spectra of the \(^{241}\hbox {Am}\) + \(^{133}\hbox {Ba}\) + \(^{137}\hbox {Cs}\) + \(^{60}\hbox {Co}\) (left) and pulse height (right) of gamma-ray signals as a function of energy at an applied voltage of 54.5 V for \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II. If no error bar is shown, the error is contained within the point.

Figure 4 shows a comparison of pulse height spectrum for \(^{133}\hbox {Ba}\) and \(^{137}\hbox {Cs}\) with \(\hbox {LaBr}_3\) (Ce) and MAPD-3NM array (I – black (75 V) and II – red (55.8 V)). We can see all gamma lines of \(^{133}\hbox {Ba}\) radioisotope in the spectrum. All photopeaks in the energy range 30 - 456 keV are well separated (Fig. 4 left). The corresponding resolution of the detector varies from 7 to 29% depending on the energy of gamma-ray and PDE of the MAPD - 3NM array.

Comparison of normalized pulse height spectrum for \(^{133}\hbox {Ba}\) (left) and \(^{137}\hbox {Cs}\) (right) with \(\hbox {LaBr}_3\) (Ce) and MAPD-3NM array (I—black and II—red).

The energy resolutions of 662 keV gamma-rays measured by MAPD-3NM-I and 3NM-II array were 4.25% and 3.3% (Fig. 4 right). MAPD-3NM-II demonstrated \(\sim\) 22% better energy resolution compared to the MAPD-3NM-I21. This improvement was due to the high PDE (33%) of the MAPD-3NM-II23.

Variation of energy resolution for gamma-ray energies of 26, 32, 56.9, 276.39, 302.85, 356, 383.84, 662, 1173, and 1332 keV were depicted in Fig. 5. The energy resolution of the detectors varied in the range of 3 - 13.6% for MAPD-3 NM-I array (59.6 - 1332 keV, at 75 V) and 2.8 - 35% MAPD-3NM-II array (30 - 1332 keV, at 54.5 V).

The energy resolution of gamma-ray signals as a function of energy. If no error bar is shown, the error is contained within the point.

In Fig. 6 shows a comparison of the energy spectra for \(^{133}\hbox {Ba}\) and \(^{137}\hbox {Cs}\) sources acquired with three different gamma detectors: commercial coaxial HPGe detector and identiFINDER-2 device (green and blue spectra, respectively) and the new developed \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II (red spectrum). The HPGe model GX3020 was manufactured by CANBERRA, while the identiFINDER-2, composed by a NaI(Tl) crystal (size Ø31 mm \(\times\) 51 mm) coupled to a PMT, was manufactured by ICx Technologies, Inc. The HPGe detector is used to clearly identified the position of gamma-rays emitted by \(^{133}\hbox {Ba}\) (Table 3). We note that there are several gamma lines associated with sum peaks (81 keV + 30 keV = 111 keV, 356 keV + 81 keV = 437 keV, 384 keV + 81 keV = 465 keV). When NaI(Tl) + PMT and \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II detectors were used, the gamma-ray spectra of radioactive sources were taken without lead shielding assemblies, therefore the natural background affected the spectra. It can be seen how the \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II is able to resolve virtually all the gamma-rays emitted by \(^{133}\hbox {Ba}\) . The non-linearity of \(\hbox {LaBr}_3\) (Ce) scintillator is visible in the lower energy region of \(^{133}\hbox {Ba}\) spectrum. In Fig. 6 is shown the spectrum of \(^{133}\hbox {Ba}\) with NaI(Tl) + PMT, too. The poor resolution of the NaI(Tl) + PMT detector does not allow separation of the gamma-rays emitted by \(^{133}\hbox {Ba}\) . As for the energy spectra of \(^{137}\hbox {Cs}\) ( Fig. 6 right part), the Compton edge and backscatter peak of gamma-rays emitted for 662 keV gamma-ray is clearly visible. The energy resolutions of 662 keV gamma-rays measured by NaI(Tl) + PMT and \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM were 8% and 3.3%, respectively. As is expected due to the improved relative energy resolution of the \(\hbox {LaBr}_3\) (Ce) than NaI(Tl) crystal, the new \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM shows a \(\sim\) 2.4 times better energy resolution than the NaI(Tl) + PMT. The parameters of the detectors used for comparison are given in Table 4.

Comparison of normalized pulse height spectrum for \(^{133}\hbox {Ba}\) (left) and \(^{137}\hbox {Cs}\) (right) with HPGe, NaI(Tl) + PMT and \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II (55.8 V) detectors.

In the case of measurements of \(^{60}\) Co with MAPD-3NM-II has applied 54.5 V voltage, due to avoid the saturation of the ADC output of the SPECTRIG MAPD module. Reducing the voltage will reduce the photon detection efficiency of the MAPD-3NM-II and degrade the energy resolution of the detectors. \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II (54.5 V) demonstrated \(\sim\) 1.8 times better energy (for 1173.2 keV and 1332.5 keV) resolution compared to the NaI(Tl) + PMT (Fig. 7).

Comparison of normalized pulse height spectrum for \(^{60}\) Co with NaI(Tl) + PMT and \(\hbox {LaBr}_3\) (Ce) + MAPD-3NM-II (54.5 V) detectors.

Obtained results proved that the newly designed MAPD-3NM-II array demonstrated good energy resolution and linearity in the wide energy region. The maximum detectable energy was 1332 keV and the number of fired pixels was about 21000 in this case (\(\sim\) 40 times smaller than the total number of pixels of the array). The number of fired pixels was calculated by taking into account PDE (30% for 380 nm) and geometrical factor of MAPD-3NM array (76% for 4 \(\times\) 4 array). It means that MAPD-3NM-II with \(\hbox {LaBr}_3\) (Ce) scintillators can be successfully used to detect tens MeV gamma-rays too. The maximum value of the recorded energy was limited due to the saturation of the ADC output in this work. In the future, it is planned to develop the following parameters of the SPECTRIG MAPD module: extending the dynamic range of the ADC up to 2 V, adding variable attenuator to the output of the MAPD, and adding a temperature compensation circuit to keep the MAPD operating with a constant gain.

The gamma-ray detection performance of the new MAPD-3NM-II array with \(\hbox {LaBr}_3\) scintillator has been studied in the energy range from 26 keV up to 1330 keV. The \(\hbox {LaBr}_3\) :Ce scintillator coupled to MAPD-3NM-II achieves an energy resolution of 3.3% at 662 keV and 4.5% for MAPD-3NM-I. The better energy resolution was related to the high PDE of the new MAPD-3NM-II photodiodes. The \(\hbox {LaBr}_3\) + MAPD-3NM-II unit delivers 3.3% more energy resolution at 662 keV than the Nal(Tl) + PMT unit, which only gives 8%. So, in other words, \(\hbox {LaBr}_3\) provides better resolution performance over NaI(Tl) systems by approximately a factor of 2.4. Note that neither the NaI(Tl) detectors nor the lanthanum bromide detectors can approach the resolution of the HPGe detector due to its low band gap of 0.67 eV.

Obtained results showed that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD. For applications such as nuclear physics, public security, industry, space application and environmental radiation monitoring, \(\hbox {LaBr}_3\) + MAPD-3NM-II unit ability to detect gamma-rays with high sensitivity is extremely important. These benefits make MAPD-3NM-II an alternative photodiode to existing instruments. In addition, MAPD-3NM-II based scintillation detectors can be considered to be more benign, in the sense that they can operate at room temperature, do not require such high voltages, increasing simplicity and a marked reduction in mass. Very small sensitive regions (4 \(\mu\) m) make these photodiodes less susceptible to radiation damage. All these advantages of the new MAPD-3NM-II makes them optimal for radiation monitoring devices that can be used in drone technology.

The datasets obtained during the experiments, used and analyzed during this study are included in this published article and its supplementary information files. The datasets are also available from the corresponding author upon reasonable request.

T. Latici, Civil and military drones, European Parliament, PE 642.230. https://www.europarl.europa.eu/RegData/etudes/BRIE/2019/642230/EPRS_BRI(2019)642230_EN.pdf (2019).

Hassanalian, M., Rice, D. & Abdelkefi, A. Evolution of space drones for planetary exploration: A review. Prog. Aerosp. Sci. 97, 61. https://doi.org/10.1016/j.paerosci.2018.01.003 (2018).

Shahmoradi, J., Talebi, E., Roghanchi, P. & Hassanalian, M. A comprehensive review of applications of drone technology in the mining industry. Drones 4, 34. https://doi.org/10.3390/drones4030034 (2020).

Pinto, L. R. et al. Radiological scouting, monitoring and inspection using drones. Sensors 21, 3143. https://doi.org/10.3390/s21093143 (2021).

ADS  Article  PubMed  PubMed Central  Google Scholar 

Connor, D. T. et al. Radiological mapping of post-disaster nuclear environments using fixed-wing unmanned aerial systems: A study from Chernobyl. Front. Robot. AIhttps://doi.org/10.3389/frobt.2019.00149 (2020).

Article  PubMed  PubMed Central  Google Scholar 

Vargas, A. et al. Comparison of airborne radiation detectors carried by rotary-wing unmanned aerial systems. Radiat. Meas. 145, 106595. https://doi.org/10.1016/j.radmeas.2021.106595 (2021).

Sasaki, M., Ishizaki, A. & Sanada, Y. Development of analysis method for airborne radiation monitoring using the inverse problem solutions. Progr. Nucl. Sci. Technol. 6, 63. https://doi.org/10.15669/pnst.6.63 (2019).

Molnar, A., Domozi, Z. & Lovas, I. Drone-based gamma radiation dose distribution survey with a discrete measurement point procedure. Sensors 21, 4930. https://doi.org/10.3390/s21144930 (2021).

ADS  Article  PubMed  PubMed Central  Google Scholar 

Sadigov, A. et al. A new detector concept for silicon photomultipliers. Nucl. Instrum. Methods A 824, 135. https://doi.org/10.1016/j.nima.2015.11.013 (2016).

ADS  CAS  Article  Google Scholar 

Sadygov, Z. et al. A new method to improve multiplication factor in micro-pixel avalanche photodiodes with high pixel density. Nucl. Instrum. Methods A 824, 137. https://doi.org/10.1016/j.nima.2015.11.008 (2016).

ADS  CAS  Article  Google Scholar 

Sadygov, Z. et al. Technology of manufacturing micropixel avalanche photodiodes and a compact matrix on their basis. Phys. Part. Nucl. Lett. 10, 780. https://doi.org/10.1134/S154747711401018X (2013).

Sadigov, A. et al. A micropixel avalanche phototransistor for time of flight measurements. Nucl. Instrum. Methods A 845, 621. https://doi.org/10.1016/j.nima.2016.06.081 (2017).

ADS  CAS  Article  Google Scholar 

Nuriyev, S. et al. Performance of a new generation of micropixel avalanche photodiodes with high pixel density and high photon detection efficiency. Nucl. Instrum. Methods A 912, 320. https://doi.org/10.1016/j.nima.2017.12.006 (2018).

ADS  CAS  Article  Google Scholar 

Nuruyev, S. et al. Performance of silicon photomultipliers at low temperature. J. Instrum. 15, C03003. https://doi.org/10.1088/1748-0221/15/03/C03003 (2020).

Akberov, R. et al. Scintillation light detection with MAPD-3NK and MPPC-S12572-010P readout. KnE Energy 3, 357. https://doi.org/10.18502/ken.v3i1.1767 (2018).

Akbarov, R. et al. Fast neutron detectors with silicon photomultiplier readouts. Nucl. Instrum. Methods A 936, 549. https://doi.org/10.1016/j.nima.2018.11.089 (2019).

ADS  CAS  Article  Google Scholar 

Ahmadov, F. et al. New gamma detector modules based on micropixel avalanche photodiode. J. Instrum. 12, C01003. https://doi.org/10.1088/1748-0221/12/01/C01003 (2017).

Akbarov, R. et al. Scintillation readout with MAPD array for gamma spectrometer. J. Instrum. 15, C01001. https://doi.org/10.1088/1748-0221/15/01/C01001 (2020).

Ahmadov, F. et al. New phoswich detector based on LFS and p-terphenyl scintillators coupled to micro pixel avalanche photodiode. Funct. Mater. 24, 344. https://doi.org/10.15407/fm24.02.341 (2017).

Ahmadov, F. et al. Alpha particle detector based on micropixel avalanche photodiodes. Phys. Part. Nucl. Lett. 10, 778. https://doi.org/10.1134/S1547477114010038 (2013).

Ahmadov, G. et al. Gamma-ray spectroscopy with MAPD array in the readout of LaBr 3: Ce scintillator. J. Instrum. 16, P07020. https://doi.org/10.1088/1748-0221/16/07/P07020 (2021).

Holik, M. et al. Miniaturized read-out interface “Spectrig MAPD’’ dedicated for silicon photomultipliers. Nucl. Instrum. Methods A 978, 164440. https://doi.org/10.1016/j.nima.2020.164440 (2020).

Ahmadov, F. et al. Investigation of parameters of new MAPD-3NM silicon photomultipliers. J. Instrum. 17, C01001. https://doi.org/10.1088/1748-0221/17/01/C01001 (2022).

Fazzi, A., Nocente, M., Tardocchi, M. et al., A large area SiPM array coupled to a LaBr3 crystal for a TPR spectrometer. in 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC). https://doi.org/10.1109/NSSMIC.2013.6829683.

Cozzi, G. et al. High-resolution gamma-ray spectroscopy with a SiPM-based detection module for 1’’ and 2’’ LaBr 3: Ce readout. IEEE Trans. Nucl. Sci. 65, 645. https://doi.org/10.1109/TNS.2017.2784238 (2018).

ADS  CAS  Article  Google Scholar 

Montagnani, G. L., Buonanno, L., Di Vita, D. & Carminati, M. Spectroscopic performance of a Sr co-doped 3’’ LaBr 3 scintillator read by a SiPM array. Nucl. Instrum. Methods A 931, 158. https://doi.org/10.1016/j.nima.2019.03.067 (2019).

ADS  CAS  Article  Google Scholar 

Marques, L., Vale, A. & Vaz, P. State-of-the-art mobile radiation detection systems for different scenarios. Sensors 21, 1051. https://doi.org/10.3390/s21041051 (2021).

ADS  Article  PubMed  PubMed Central  Google Scholar 

https://www.epic-crystal.com/halide-scintillators/cebr3-scintillator.html

This work was supported by the Scientific Foundation of SOCAR. This work was also supported by OP VVV Project CZ.02.1.01/0.0/0.3/16-019/0000766 (Engineering applications of microworld physics) and LTT17018 (Research infrastructure CERN).

Faculty of Electrical Engineering, University of West Bohemia, Univerzitni 2795/26, 301 00, Pilsen, Czech Republic

M. Holik, O. Urban & J. Zich

Institute of Experimental and Applied Physics, Czech Technical University in Prague, Husova 240/5, 110 00, Prague, Czech Republic

M. Holik, F. Mamedov & I. Stekl

Institute of Radiation Problems-ANAS, B. Vahabzade Str. 9, AZ1143, Baku, Azerbaijan

F. Ahmadov, A. Sadigov, G. Ahmadov, Z. Sadygov & A. Garibli

National Nuclear Research Center Under the MDDT, Baku Shamakhy HW 20 km, Gobu Sett. of Absheron Dist, AZ0100, Baku, Azerbaijan

F. Ahmadov, A. Sadigov, G. Ahmadov, J. Naghiyev, Z. Sadygov, A. Mammadli & A. Garibli

The Institute of Nuclear Physics, Ibragimova 1, 050032, Almaty, Kazakhstan

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

M.H., A.S., F.A., G.A. did experiments and wrote the main manuscript text. Z.S., I.S. - Planning of experiments. D.B., F.M., J.N., A.M. and A.G. - Processing and evaluation of measured data. O.U., J.Z. - Reviewed the manuscript.

Correspondence to M. Holik or F. Ahmadov.

The authors declare no competing interests.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Holik, M., Ahmadov, F., Sadigov, A. et al. Gamma ray detection performance of newly developed MAPD-3NM-II photosensor with LaBr3(Ce) crystal. Sci Rep 12, 15855 (2022). https://doi.org/10.1038/s41598-022-20006-z

DOI: https://doi.org/10.1038/s41598-022-20006-z

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.

Scientific Reports (Sci Rep) ISSN 2045-2322 (online)

Sign up for the Nature Briefing newsletter — what matters in science, free to your inbox daily.